Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

FMB50 static pressure level gauge

FMB50 static pressure level gauge

Deltapilot M FMB50

Configurable Models 1 models
Integrated pressure sensor using Contite measuring elements for static pressure liquid level measurement
Compare

Continuous / Liquids

Measuring Principle Hydrostatic
Characteristic / Application Compact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant and climatic-proofed
Specialities Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions
Hermetically sealed Contite measuring cell (condensate-resistant and climatic-proofed)
Supply / Communication 4...20mA HART:
10,5...45V DC
Ex ia: 10,5...30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9...32V DC (Non Ex)
Accuracy Standard 0.2%
Optional 0.1%
Long term stability 0.05% of URL/year
Ambient temperature -40°C...85°C
(-40°F...185°F)
Process temperature -10°C...100°C
(14°F...212°F)
Process pressure / max. overpressure limit 40 bar (600 psi)
Pressure measuring range 100 mbar...10 bar
(1.5 psi...150 psi)
Main wetted parts Alloy C
316L  
optional coating AuPt
Process connection Thread:
G1 1/2, MNPT1 1/2
Flange:
DN40...DN100,
ASME 1 1/2"...4",
JIS 10K
Process connection hygienic Tri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent N
SMS
DRD
Max. measurement distance 100m (328ft) H2O
Communication 4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
IO-Link
Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI
Safety approvals Overfill protection WHG
SIL
Design approvals EN10204-3.1
NACE MR0103
Hygienic approvals 3A, EHEDG
CoC ASME-BPE
Marine approval GL/ ABS/ LR/ BV/ DNV
Drinking water approvals KTW/ NSF/ ACS
Options Separate housing  
Initial device settings
Overvoltage protection
Application limits If pressurized, usage of two pressure transmitters to measure the differential pressure (electronic dP)
Observe ratio head pressure : hydrostatic pressure

Pressure

Measuring Principle Hydrostatic pressure
Characteristic Compact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences
Supply voltage 4...20 mA HART
10,5...45 VDC (Non Ex):
Ex ia: 10,5...30 VDC
PROFIBUS PA:
9...32 VDC (Non Ex)
FOUNDATION Fieldbus:
9...32 VDC (Non Ex)
Reference Accuracy Standard 0.2 %
Optional 0.1 %
Long term stability 0.05 % of URL/ year
Process temperature -10°C...100°C
(14°F...212°F)
Ambient temperature -40°C...85°C
(-40°F...185°F)
Measuring cell 100 mbar...10 bar
(1.5 psi...150 psi)
Smallest calibratable span 10 mbar (1.45 psi)
Vacuum resistance 0 mbar abs.
Max. Turn down 100:1
Max. overpressure limit 40 bar (600 psi)
Process connection Thread:
G1 1/2, MNPT1 1/2
Flange:
DN40...DN100,
ASME 1 1/2"...4",
JIS 10K
Process connection hygienic Tri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent N
SMS
DRD
Material process membrane 316L, AlloyC
Gold-Rhodium
Material gasket None, measuring cell welded
Fill fluid Inert oil,
Synthetic oil
Material housing 316L, Die-cast aluminum
Communication 4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
IO-Link
Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
Safety approvals Overfill protection WHG
SIL
Design approvals EN10204-3.1
NACE MR0103
Hygienic approvals 3A, EHEDG
CoC ASME-BPE
Marine approvals GL/ ABS/ LR/ BV/ DNV
Drinking water approvals KTW/ NSF/ ACS
Specialities Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions

The Deltapilot FMB50 compact transmitter uses the condensation-resistant Contite measuring cell and is mainly used in the process industry and the hygiene sector to measure the level of liquid or paste-like media in open or closed vessels. Measurement is unaffected by foam and meets SIL2 functional safety requirements.

 

Advantages

Excellent repeatability and long-term stability

Turndown ratio: 100:1

Fully sealed CONTITE measuring cell: condensation-resistant, weather-protected; high reference accuracy: ±0.2%, optional: ±0.1%; minimal temperature influence

Point-to-point modules for measuring differential pressure, static pressure and pressure (Deltabar M - Deltapilot M - Cerabar M), e.g. replaceable display, universal electronics

Simple commissioning without operating tools

 

Areas of application

Hydrostatic sensors can be used for the following measurements:

 

Hydrostatic measurement of liquid and paste-like media in the process industry, process measurement, pharmaceutical and food industries

Measurement of liquid level, volume or mass

 

 

 

E+H Product Configurator

Please enter the product order code to configure

Product comparison

Please enter the order codes of the products to compare in the boxes below (up to three products)

 

   

Technical documentation

   

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

Source:Endress+Hauserwebsite public information

Model Name
FMB50-BA21RA1PGBCFJB3ULA 静压液位计
Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · Static Pressure Measurement Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of Static Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Static Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · Static Pressure Measurement Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of Static Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Static Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · Static Pressure Measurement Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of Static Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Static Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · Static Pressure Measurement Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of Static Pressure Measurement.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Static Pressure Measurement for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.