Thermoreactors for easy digestion of standard parameters
Safe and easy operation with automatic heating and switch-off for standard digestions
• 12 round cells (16 mm)
• 8 pre-programmed settings including fast digestion for COD
• Easy, quick operation for standard parameters
The required temperatures and degradation times for the standard parameters are stored in every WTW thermoreactor. In addition, there are different options for self programming and cuvette numbers available.
CR 2200 allows standardi digestion with programs for COD, total nitrogen, total phosphorus, TOC and heavy metals of galvanizing. The high reaction temperature over a defined period of time ensures a complete digestion of the sample. For COD, an additional program for fast digestion (148°C, 20 min) is available.
Programs for standard parameters
•1: 148 °C for 120 minutes
•2: 120 °C for 30 minutes
•3: 120 °C for 60 minutes
•4: 120 °C for 120 minutes
•5: 100 °C for 60 minutes
•6: 148 °C for 20 minutes
•7: 150 °C for 120 minutes
•8: 100 °C for 30 minutesReactor type Dry temperature control device with safety cover Cell shafts 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm Reaction time setting program settings or user-defined programs up to 180 min Temperature setting stored program, user-defined settings from room temperature up to 170°C Controlling accuracy ± 1 °C ± 1 digit Temperature stability ± 0.5 K Overtemperature protection 190 °C ± 5 °C Dimensions D x W x H: 312 x 255 x 185 mm Weight: 4 kg Output unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC Power supply 115/230 VAC 50/60 Hz switchable via fuse setting
allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 ATProtection Class IP 20 according to EN 60529 Ambient temperature Operation +5 °C to +40 °C Climatic class 2 according to VDI/VDE 3540 Relative humidity: Yearly mean: < 75 % 30 days /year: 95 % Other days: 85 % Light dew: yes Site altitude: max. 5000 m above sea level EMC EN 61326-1 FCC 47 CFR Part 15 Test certificates cETLus, CE Safety standards EN 61010-1 EN 61010-2-010 UL 61010-1 CAN/CSA C22.2#61010-1 CAN/CSA C22.2#61010-2-010
1P21- CR 2200
用于COD和热消化的热反应器,最多可容纳12个反应杯,有4个反应器温度,8个固定程序,115/230 V,区域插头。