Turb750IR

600120 Turb 750 IR

Laboratory turbidimeter with infrared LED according to DIN EN ISO 7027-1, universal power supply 90 to 240 VAC, single instrument with calibration standard kit (0.02 - 10.0 - 1000 NTU), 3 empty vials, Pharmcopeia compliant, optional 1-5 point calibration, quick cal, USB and RS232 Interface, LSData PC software for GLP-compliant Data management.
符合DIN EN ISO 7027-1的红外LED实验室浊度计,通用电源90至240 VAC,带校准标准套件(0.02-10.0-1000 NTU)的单台仪器,3个空瓶,符合Pharmacopeia标准,可选1-5点校准,快速校准,USB和RS232接口,用于GLP标准数据管理的LSData PC软件。
售价:
¥ 0.00
原价: ¥0.00
关注:0

服务:  现货现款・ 48小时快速处理・ 含税运。

收藏
数量:
  • WTW Turb® 750 IR fulfills the demands of many lab and industrial applications for a AQA supported turbidity measurement up to 1100 FNU/NTU. With the Intelligent Reproducibility and Plausibility-Check (IRPC) reliable findings are proven.

    Measurement conforming to standards in drinking water, lab and industrial applications.

    WTW Turb® 750 IR measures nephelometrically according to DIN EN ISO 7027-1 up to 1100 NTU/FNU with AQA support and PC documentation:

    • Well proven WTW® optics with fast data output
    • Multiple measurement and data evaluation with Intelligent Reproducibility and Plausibility Check (IRPC)
    • Bench-saving despite comprehensive functionality
    • Invincible price-/performance ratio

    Turbidity stands for an indicator parameter on quality in drinking water, waste water process and many industrial applications. The measurement is support by:

    • Measurement procedure fulfilling DIN EN ISO 7027-1 *)
    • Function keys
    • Comfortable menu-guided 3-point standard calibration
    • Flexible calibration with 2-5 user-defined calibration points
    • QuickCAL: 1-point-calibration at 10 NTU
    • AQA, calibration interval and records with PC-documentation

    *) For drinking water measurement acc. to  US EPA 180.1 Turb® 750 T is the appropriate instrument.

    More informationen in: "Turbidity An interesting parameter - Brief Overview, Measurement Tips & Tricks" and
    in our video: "https://www.youtube.com/watch?v=UW4Mk0IrOww"

     

    • Light source  LED  830-890 nm acc. to DIN EN ISO 7027-1
      acc. to standard methods DIN EN ISO 7027-1, DIN 38404
      Modes Nephelometric (scattering 90°)
      Display Backlit graphics display 160x104
      Keypad Foil keypad with tactile feedback alphanumeric entry option
      Range 0-1100 FNU/NTU
      Units FNU/NTU
      Resolution (Measurement resolution max. 0,02 NTU due to natural scattering of water; digital resolution >0,02 if optional ) 0,01 for 0,00 -9,99, 0,1 for 10 – 99,90, 1 for 100 - 1100
      Accuracy NTU/FNU 0.01 NTU/FNU or +/- 2% of finding
      Repeatability < 0.5% of finding
      Reading modes Findings with Intelligent Reproducibility and Plausibility Check (IRPC),
      rapid settlement samples supported by quick response time and IRPC
      Response time 4 sec 
      Calibration - options Automatic 3-P Kal, (Default),  2-5 P user calibration, QUICKCal
      Calibration protocol and interval setting yes / yes
      AQS-Support Calibration protocol and interval setting, AQA flag, Cal flag 
      Data storage 2500 data sets with Cal protocol, AQA flag, GLP comliant data management with LIMS-connection
      Sample Identification yes - alphanumeric 
      Air Purge no
      Firmware-Update yes via USB
      Interface RS 232, USB, Printer via PC or RS232
      Storage condition instrument  - 25 °C ... + 65 °C (13-149 °F)
      Operating Temperature Range +5 °C ... +55 °C (41-131 °F) with power plug connected: +5 °C ... +40 °C  (41-104 °F)
      Data management GLP-compliant incl. PC software USB, RS232 Turb® Data
      Dimensions (H x W x D) ca. 290 x 190 x 80 mm
      Weight 1,1 kg
      Calibration Standards Longterm stable Cal.Kits for standard 3-P calibration; 0,02-10-1000; FNU/NTU; Polymer AMCO® Clear standards: longterm stability, higly precise, optimized to optics, production accuracy ± 1%,  non-hazardous 
      Vials , Sample Volume Vials 28 mm diameter, volume min. 15 ml; Borosilicate, cap: phenolic resin, PTFE coated sealing gasket.; NO SILICON OIL to cover scratches required for measurement procedure!
      Sample conditions sample temperature < 70°C (158°F)
      Power supply Wide range power supply with plugs for Euro, US, UK and Australia
      Certificates CE
      Delivery Lab turbidimeer Turb® 750 IR; 4 batteries 1,5 V Typ AA; Wide range power supply; cable USB-A to USB-B mini; 6 empty vials, sticker for orientation marking; Cal.Kit Turb® 430/750 IR; Cloth, quick guide, compact operation manual, CDROM with extended manual, CDROM with PC software Turb® Data; Inspection protocol

     

      •  