1P22- CR 3200
用于COD和热消化的热反应器,最多24个单元,一个程序设置,8个固定/8个用户定义程序,自由温度设置高达170°C,230/115V,区域插头。可选配件:用于AQA的外部温度传感器TFK CR-温度监测(250 100)。
-
-
Digestions of 24 round cells with pre-programmed and user-defined programs
In total 24 round cells (16 mm) in two seperate heating blocks using the same program
• 8 stored programs for easy and fast operation including COD fast digestion
• 8 user-defined programs from room temperature/25-170°C with high temperature digestion
• AQA/IQC with optional external temperature sensor for documentation
Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.
The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.
Programs for standard parameters
•1: 148 °C for 120 minutes
•2: 120 °C for 30 minutes
•3: 120 °C for 60 minutes
•4: 120 °C for 120 minutes
•5: 100 °C for 60 minutes
•6: 148 °C for 20 minutes
•7: 150 °C for 120 minutes
•8: 100 °C for 30 minutesReactor type Dry temperature control device with safety cover Cell shafts 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm Reaction time setting program settings or user-defined programs up to 180 min Temperature setting stored program, user-defined settings from room temperature up to 170°C Controlling accuracy ± 1 °C ± 1 digit Temperature stability ± 0.5 K Overtemperature protection 190 °C ± 5 °C Dimensions D x W x H: 312 x 255 x 185 mm Weight: 4 kg Output unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC Power supply 115/230 VAC 50/60 Hz switchable via fuse setting
allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 ATProtection Class IP 20 according to EN 60529 Ambient temperature Operation +5 °C to +40 °C Climatic class 2 according to VDI/VDE 3540 Relative humidity: Yearly mean: < 75 % 30 days /year: 95 % Other days: 85 % Light dew: yes Site altitude: max. 5000 m above sea level EMC EN 61326-1 FCC 47 CFR Part 15 Test certificates cETLus, CE Safety standards EN 61010-1 EN 61010-2-010 UL 61010-1 CAN/CSA C22.2#61010-1 CAN/CSA C22.2#61010-2-010
-