Accuracy ± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂
± 10% from 5 - 10 mg/L Cl₂Air Purge Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar (20 psig) maximum Alerts Low and high chlorine. Dirty cell. Low and high sample flow. Certifications CE compliant with: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
South Korea KC Certificate
North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003Controller Compatibility SC200, SC1000, SC4500. All controllers sold separately Cycle Time 2.5 minutes Dimensions (H x W x D) 342 mm x 329 mm x 177 mm Drain ½-inch ID flexible hose Enclosure Rating IP66 Inlet ¼-inch OD polyethylene tube, quick-disconnect fitting Inlet Pressure 0.3 - 5.2 bar (4.5 - 75 psig) supplied to Y-strainer
0.1 - 0.3 bar (1.5 - 5 psig) supplied to analyzerInstallation Kit Pressure Regulator Installation Kit Interferences Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃. Light Source LED, measurement at 510 nm, 1 cm light pathlength Lower Limit of Detection (LOD) 0.03 mg/L Mounting Wall mount Operating Humidity 0 - 90% relative humidity, non-condensing Operating Temperature Range 5 - 40 °C (41 - 104 °F) Parameter Chlorine, free; Chlorine, total residual Power Requirements (Voltage) 12 VDC, 400 mA maximum (supplied by the controller) Range 0 -10 mg/L Cl₂ Reagent Consumption 0.5 L of buffer solution and 0.5 L of indicator solution in 31 days Sample Flow Rate 60 - 200 mL/min through the instrument Sample Requirements Y-strainer filtration with 40-mesh screen or higher Sample Temperature 5 - 40 °C (41 - 104 °F) Warranty 12 months Weight 4.1 kg (9.0 lbs.) What's included? CL17sc Chlorine Analyser with Pressure Regulator Installation Kit, and User Manual. User provides screws and/or anchors for mounting.
Hach SC Controller and Reagents are not included, they are sold separately.